10. Is it possible to change the scan rate on the 4 Channel Controller?

Yes. The default scan rate is approximately 3 seconds and this is suitable for most applications.

However, the Scan Rate is user programmable from System Setup mode.

Press the CAL button and the STATUS display will indicate CA. Press the SET button until the STATUS display indicates CS for "Channel Scan". The VALUE display will indicate the current scan time in 100mS intervals. Press the up or down arrow buttons until the STATUS display indicates the required value. Press the CAL button to save the new value. Futher details on changing the scan rate are available in section 3 of the instruction manual.